㩷 Control specifications
1. Keys on the control panel
BUZZER : In alarm condition, buzzer stops sounding with this key pressed.
Remote alarm output and alarm message would not be off.
During the power failure (battery back-up), press the key to show the
chamber temperature for 5 seconds.
ALARM TEST : With this key pressed to activate alarm test mode to be forcibly step into
alarm condition (ALARM lamp blinks, intermittent buzzer beeps, digital
LED goes off and remote alarm activates).
After approx. 90seconds then, unit retrieves normal condition by Auto
Return.
With the battery switch turns off, “E09” blinks on the display.
SET : Press this key once to blink 2
nd
digit on the display and press the key
again to store the value to be changed.
During setting mode, press the key to shift the blinking digit.
In chamber temperature display, press this key over 5 seconds to shift to
(Digit shift key)
Key Lock screen with ‘L_0’ displayed.
During setting mode, press the key to change the blinking digit. In
chamber temperature display, press the key over 5 seconds to enter the
(Numerical value
function mode. (“F00” is displayed)
shift key)
2. Temperature control
Setting range :
-50㷄~-90㷄
Display range :
-180㷄~+50㷄
Setting procedure : Press SET key and set the required value with key and key.
Press SET key to store the set value.
Unacceptable setting If you try to set the value which deviates from setting range, the buzzer
range : beeps for 1 second to inform the value unacceptable.
3. Key Lock mode
Setting range : 0, 1
Setting procedure: In chamber temperature display, keep pressing
key over 5 seconds
to enter into Key Lock mode. (“L_0” or “L_1” is displayed. initial: L_0)
with “0” or “1” blinks.
Change the value with
key and press SET key to store the value in
the non-volatile memory.
0: Unlocked
1: Locked
4. Function mode
Setting range : 00~32
Display range : 00~39
00, 16 and 33~39 are unavailable.
Setting procedure : In chamber temperature display, keep pressing key over 5 seconds
to enter function mode (F00 is displayed). Change the blinking 1
st
digit
to desired function code with
key and key. Press SET key to be
function code available.
If you input 00 and 16 then press SET key, automatically revert to
chamber temperature display.
Unacceptable setting If you try to input the value among F33~F39, the buzzer beeps to
range : inform the value unacceptable.