7. PARAMETER SETTING VM-5S
-52-
7-9 AUTO TRIGGER MODE SELECTION
(1) Remove the monitor unit from the rack.
• Refer to the instruction manual of the instrument rack in which the monitor is
installed.
(2) The Trigger Mode switch is located on the trigger card on the main board. Set the switch to the
left for Auto Trigger or to the right for Manual Trigger.
• Do not touch any parts not specifically mentioned.
(3) After verifying the trigger mode selected, reinstall the monitor unit in the rack.
• Refer to the instruction manual of the instrument rack in which the monitor is
installed.
Mandatory
Mandatory
VM-5 TRIGGER CARD
J1
CH1
HYS. SET
0.1V
0.2V
0.5V
1.0V
CH1 TRG. MODE
AUTO MAN.
VM-5
TRIGGER
CARD
J3
CH2
HYS. SET
0.1V
0.2V
0.5V
1.0V
CH2 TRG. MODE
AUTO MAN.
S1
S2
Mandatory