Performance Check Procedure
c. Set leveled sine-wave generator output to produce a 7.0 division sine-
wave display amplitude at 500 kHz.
d. Add a 10X and a 2X BNC attenuator before the 50 f l BNC termination
(for a 0.35 division display amplitude).
e. CHECK—that the display cannot be triggered In either HF REJ of
NOISE REJ CPLG.
f. Set:
Horizontal MODE B
A/B SELECT B Trigger
B SEC/DIV 1 |1S
g. CHECK—that the display cannot be triggered in HF REJ or NOISE REJ
CPLG by adjusting the Trigger LEVEL control.
3. 25 MHz Trigger Sensitivity
a. Set:
A
A Trigger
DC
50 ns
Horizontal MODE
A/B SELECT
TRIGGER CPLG
A SEC/DIV
b. Remove the 10X and 2X BNC attenuators from the signal path.
c. Set leveled sine-wave generator output to produce a 7.0 division dis
play amplitude at 25 MHz.
d. Add a 10X and a 2X BNC attenuator before the 50 f l BNC termination.
e. CHECK—that the display Is stably triggered In DC, LF REJ, and AC
Trigger CPLG; the display is not triggered in NOISE REJ and HF REJ
Trigger CPLG settings.
f. Set:
TRIGGER CPLG
Horizontal MODE
A/B SELECT
B SEC/DIV
DC
B
B Trigger
20 ns
7-30 2245A Operators

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2245A and is the answer not in the manual?

Tektronix 2245A Specifications

General IconGeneral
BrandTektronix
Model2245A
CategoryTest Equipment
LanguageEnglish

Related product manuals