EasyManuals Logo
Home>Tektronix>Test Equipment>2245A

Tektronix 2245A User Manual

Tektronix 2245A
194 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #159 background imageLoading...
Page #159 background image
Performance Check Procedure
e. Set A INTEN to midrange.
f. Disconnect the test equipment from the 2245A.
2. PROBE ADJUST O utput
a. Set:
CH 1 VOLTS/DIV 10 mV
SEC/DIV 0.2 ms
b. Connect a 10X probe to the CH 1 input connector and connect the
probe tip to the 2245A PROBE ADJUST output. (When using Tektronix
coded probes, the readout changes to .1 V.)
c. CHECK—For a 5-division vertical display of PROBE ADJUST square-
wave signal (square-wave period is typically 1 ms, within 25%).
3. AUTO SETUP Functional Check
a. Set:
CH 1 COUPLINQ GND
CH 1 VOLTS/DIV 2 mV
A SEC/DIV 20 ns
b. Press the AUTO SETUP button.
c. Check that the Probe Adjust waveform is stably displayed on the upper
half of the crt.
2245A Operators
7-49

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2245A and is the answer not in the manual?

Tektronix 2245A Specifications

General IconGeneral
BrandTektronix
Model2245A
CategoryTest Equipment
LanguageEnglish

Related product manuals