EasyManuals Logo
Home>Tektronix>Test Equipment>2445B

Tektronix 2445B User Manual

Tektronix 2445B
207 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #90 background imageLoading...
Page #90 background image
Checks and Adjustments
Probe Compensation
Accurate measurements require accurate probe compensation. To ensure
optimum measurement accuracy, check probe compensation any time a probe is
attached to the instrument or any other time you are not certain of correct
compensation. Because of minor differences between channels, CH 1 and CH 2
probes should be compensated on their respective channels. CH 3 and CH 4
probes should be compensated on CH 1 or CH 2. Check and adjust probe low-
frequency compensation as follows:
1.
Obtain a display as described in "Initial Setup."
2.
Set the SEC/DIV control to
1
ms and 20 MHz BW LIMIT
on.
If the probe to
be compensated is connected to CH 2, enable the Channel 2 display. Set
the appropriate VOLTS/DIV control to 100 mV.
3. Connect the probe to the CALIBRATOR output.
4.
Check the waveform for overshoot and rolloff (see Figure 4-1). If necessary,
adjust the probe for a square front corner on the waveform, using the small
adjustment tool supplied in the probe accessory package. Insert the tool
through the small hole in the side of the box attached to the vertical input
connector.
100 mV
(3832-12) 5854-19
Figure
4-1.
Probe low-frequency compensation.
4-4
2465B/2455B/2445B Operators

Table of Contents

Other manuals for Tektronix 2445B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2445B and is the answer not in the manual?

Tektronix 2445B Specifications

General IconGeneral
BrandTektronix
Model2445B
CategoryTest Equipment
LanguageEnglish

Related product manuals