Risetime: 3.5 nanoseconds
or
less (calculated from
0.35
-:-
bandwidth in MHz).
Maximum Input Voltage:
De-coupled
250 V
(de+
peak
ac)
or
500 V p-p ac at 1 kHz
or
less; ac-coupled, 500 V (de
+ peak ac)
or
500 V
p-p
ac 1 kHz
or
less.
Positive-Going Step Aberrations: Less than +3%,
-3%,
not
to
exceed
3%
peak-to-peak,
excluding
the ADD mode.
Common-Mode Rejection Ratio
(ADD
Mode with CH 2
Inverted): At least
10:1
at 20 MHz
for
common mode
signals
of
6 divisions
or
less
with
GAIN adjusted
for
best
CMRR at
50
kHz.
Input Gate Current: 0.5
nA
or
less
(0.1
div at
5 mV/DIV), from
-15°C
to
+30°C.
Channel Isolation: At least 100:1 at
25
MHz.
CHOP
Mode Repetition Rate: Approximately 250 kHz.
Cascaded Operation
(CH
1
OUT
int.o
CH
2)
sensitivity:
Approximately
1 mV/DIV.
Bandwidth
is de to at least
50
MHz,
with
CH 2
OUT
connected
to
CH 1 input. AC-
coupled, using a 50
0,
42-inch
BNC
cable, terminated in
50
n at CH 2 Input.
TRIGGERING
Sensitivity
Ac Coupled Signal: 0.3 div internal
or
50 mV external,
from 30 Hz
to
25
MHz; increasing
to
1.5 div, internal
or
150 mV external at 100 MHz.
LF
REJ
Coupled Signal: 0.5 div internal
or
100 mV
external, from
50
kHz to
25
MHz; increasing to 1.5 div
internal
or
300 mV external at 100 MHz. Attenuates
signals below about 50 kHz.
HF
REJ
Coupled Signal: 0.5 div internal
or
50 mV
external, from 30 Hz to 50 kHz. Attenuates signals below
about
50 Hz and above
about
50
kHz.
DC Coupled Signal: 0.3 div internal
or
50 mV external,
from de to
25
MHz; increasing
to
1.5 div internal
or
150 mV
external at 100 MHz.
EXT
-:-
10 Signal:
Amplitude
requirements are
mul-
tiplied
by
10.
External Trigger Input
Maximum Input Voltage: 250 V
(de+
peak ac)
or
250 V
p-p ac
(1
kHz
or
less).
66
466/464 DM44 Operators
@
I
I
I
I
I
I
I
I
I