Getting Acquain
tedwithYourInstrument
Quick Tips
Interleaving is applied to analog output. When interleaving is on, marker data with even numbers will be output, such
as 0,0,2,2,4,4,6,6.
When output is on, changing the Interleave m ode on and off will turn the output off.
The following table lists how interleaving extends the maximum sampling rate and waveform length:
Instrument type an d
option Interleave Sampling rate Waveform length
On 12 GS/s to 24 GS/s
2 to 129,600,000 points
AWG7122C, AWG7122B
(Option 01, Option 06)
Off 10MS/sto12GS/s
1 to 64,800,000 points
On 8GS/sto12GS/s
2 to 129,600,000 points
AWG7082C
(Option 01, Option 06)
Off 10MS/sto8GS/s
1 to 64,800,000 points
On 12 GS/s to 24 GS/s
2 to 64,800,000 points
AWG7122C, AWG7122B
(Option 06)
Off 10MS/sto12GS/s
1 to 32,400,000 points
On 8GS/sto16GS/s
2 to 64,800,000 points
AWG7082C
(Option 06)
Off 10MS/sto8GS/s
1 to 32,400,000 points
AWG7122C, AWG7122B
(Option 01)
Not a vailable
10 MS/s to 12 GS/s
1 to 64,800,000 points
AWG7082C
(Option 01)
Not a vailable
10 MS/s to 8 GS/s
1 to 64,800,000 points
AWG7122C, AWG7122B
(Standard models)
Not a vailable
10 MS/s to 12 GS/s
1 to 32,400,000 points
AWG7082C
(Standard model)
Not a vailable
10 MS/s to 8 GS/s
1 to 32,400,000 points
AWG5000 and AWG7000 Series Quick Start User Manual 37