Calibration Procedure
DM 501A
Performance
Check Procedure
Table
4-8
FRONT PANEL AC CURRENT ACCURACY
!
Range
'
I
I
Current
DISPLAY
LIMITS
Ambient Temp. Range
T18°C to +28“C
Ambient
Temp. Range
O' Cto
H8“C, -H28°C
to I50“C
200 //A
190.00M
191.24 to 188,76 191.48 to 188.52
2
mA 1.9000
r))A
1.9124 to 1.8876 1.9148
to
1.8852
20
mA 19.000 mA 19.124 to 18.876 19.148
to
18.852
200 rnA
190.00 mA 191.24 to 188.76
191.48
to
188.52
Pf'OO
inA 1900.0
mA
1912.4 to
1887.6 1914.8
to
1885.2
Table
4-9
TEMPERATURE
LIMITS
DISPLAY LIMITS
Temperature
Probe Calibrated To Instrument Any Probe
i
Ambient
Temp. Range
Ambient
Temp.
Range
(
18'
C to
i
28“C
0"C to
t
18"C
I
28''C
to
I
50"
C
118
C to
f28"C
1
0“Cto-M8“C
1
f-28°C to t 50
C
62"
C
GO to
64 58.5 to 65.5 58 to 66
-56.5
to
67.5
125“C
1
123
to
1127
I
121.5 to
i 128.5
!21
to
t
129
1
t
119.5 to
t
130.5
t
200"
C
1
i
194 10
1 200
i
192.5
to
: 201 5 192 to
1 202
I
i
190.5 to
i 203.5
Table
4-10
REAR INTERFACE
DC VOLTAGE
ACCURACY
Range
Dc
Calibration
Voltage
DISPLAY LIMITS
Ambient
Temp.
Range
1 18°C to
I
28'C
Ambient Temp. Range
O’C
to
f
18"C,
4
28"C
to t-50"C
200 mV
j
190,00
mV
190.
13
to
189,87
1
190 24 to 189.76
2
V
:
1,9000 V
1.9012 to
1.8988
1.9024 to
1.8976
20
V
19,000 V
19,012 to 18
988
19.024 to
18.976
200 V
190,00 V
190.12
to
189.88
190.24 to
189.76
1000 V
1 90
00 V
1
190.3 to
189.7
190.7 to 189.3
4 8