DMM91X Series Performance Verification
Handheld and Benchtop Instruments Basic Service
9
Test Equipment
The performance verification procedures use external traceable test equipment to
directly check warranted characteristics. If you substitute equipment, always
choose instruments that meet or exceed the minimum requirements listed in
Table 8.
Alternative test equipment must meet or exceed the intended minimum
requirements. If you substitute equipment, you may need to modify the
performance verification procedures.
NOTE. Before beginning the performance verification procedures, warm up the
test equipment according to the manufacturers recommendations.
Table 8: Test Equipment
Description Minimum Requirements Example Product
ÁÁÁÁÁÁÁÁ
Á
ÁÁÁÁÁÁ
Á
Á
ÁÁÁÁÁÁ
Á
ÁÁÁÁÁÁÁÁ
Universal Calibration
System
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
Resolution & accuracy 4 times
greater than the multimeter
display reading.
ÁÁÁÁÁÁÁÁÁ
Á
ÁÁÁÁÁÁÁ
Á
Á
ÁÁÁÁÁÁÁ
Á
ÁÁÁÁÁÁÁÁÁ
Wavetek 9100 with 9105 lead
set.
AC and DC volts measurement
1
AC and DC current measurement
Resistance measurement
1
Capacitance measurement
Sinewave generation
Squarewave generation
Thermocouple adapter K Type Tektronix ATK01
Capacitance Standard Optional
1
Choose 4-wire measurement setup if available.

Table of Contents

Other manuals for Tektronix DMM912

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DMM912 and is the answer not in the manual?

Tektronix DMM912 Specifications

General IconGeneral
BrandTektronix
ModelDMM912
CategoryMultimeter
LanguageEnglish

Related product manuals