PS280 and PS283 Performance Verification
6
Handheld and Benchtop Instruments Basic Service
Test Equipment
The performance verification procedures use external traceable test equipment to
directly check warranted characteristics. If you substitute equipment, always
choose instruments that meet or exceed the minimum requirements specified in
Table 6.
Alternate test equipment must meet or exceed the intended minimum require-
ments. If you substitute equipment, you may need to modify the performance
verification procedures.
NOTE. Before beginning the performance verification procedures, warm up the
test equipment according to the manufacturers recommendations.
Table 6: Performance Verification and Adjustment Test Equipment
Description Minimum Requirements Example Product
Oscilloscope 50 MHz, 3% vertical deflection accuracy Tektronix TDS300 Series Oscilloscope
Digital Multimeter 35 V, 4 A, ±0.1% Accuracy Tektronix DMM914 Digital Multimeter
Fixed Resistive Loads 20 30 W
10 30 W
2 5 W
1 30 W
7 140 W
0.5 5 W
15 70 W
30 140 W
7.5 140 W
30 40 W
60 70 W
Variable Resistive Load 0 to 2 30 W
Shorting Strap with Alligator Clips Insulated

Table of Contents

Other manuals for Tektronix DMM912

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DMM912 and is the answer not in the manual?

Tektronix DMM912 Specifications

General IconGeneral
BrandTektronix
ModelDMM912
CategoryMultimeter
LanguageEnglish

Related product manuals