Commands Listed in Alphabetical Order
CH<x>:YUNits
Sets or returns the units for the channel specied by <x>, where x is the channel
number. String arguments are case insensitive and any unsupported units will
generate an e
rror.
Supported units are:
%, /Hz, A, A/
A, A/V, A/W, A/dB, A/s, AA, AW, AdB, As, B, Hz, IRE, S/s, V,
V/A, V/V, V/W, V/dB, V/s, VV, VW, VdB, Volts, Vs, W, W/A, W/V, W/W, W/dB,
W/s,WA,WV,WW,WdB,Ws,dB,dB/A,dB/V,dB/W,dB/dB,dBA,dBV,dBW,
dBdB, day, degrees, div, hr, min, ohms, percent, s
Group
Vertical
Syntax
CH<x>:YUNits <QString>
CH<x>:YUNits?
Arguments
QString is a string of text surrounded by quotes, specifying the supported units.
CLEARMenu (No Query Form)
Clears the current menu from the display. This command is equivalent to pressing
the front panel Menu off.
Group
Miscellaneous
Syntax
CLEARMenu
*CLS (No Query Form)
Cl
ears the following:
Event Queue
Standard Event Status Register
Status Byte Register (except the MAV bit)
If the *CLS command immediately follows an <EOI>, the Output Queue and
MAV bit (Status Byte Register bit 4) are also cleared. The MAV bit indicates that
information is in the output queue. The device clear (DCL) GPIB control message
will clear the output queue and thus MAV. *CLS does not clear the output queue
o r M AV.
2-11 0 MSO2000B, DPO2000B, MSO2000 and DPO2000 Series O scilloscopes Programmer Manual

Other manuals for Tektronix DPO2012

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO2012 and is the answer not in the manual?

Tektronix DPO2012 Specifications

General IconGeneral
BrandTektronix
ModelDPO2012
CategoryTest Equipment
LanguageEnglish

Related product manuals