Commands Listed in Alphabetical Order
ACQuire? (Query Only)
Returns the f
ollowing current acquisition parameters:
Stop after
Acquisition state
Mode
Number of averages
Sampling mode
Group
Acquisition
Syntax
ACQuire?
Related Commands
ACQuire:MODe, ACQuire:NUMACq?, ACQuire:NUMAVg, ACQuire:
STOPAfter
Examples
ACQuire? might return ACQUIRE:STOPAFTER RUNSTOP;STATE 1;MODE
SAMPLE;NUMAVG 16;SAMPLINGMODE RT
ACQuire:MAXSamplerate? (Query O nly)
Returns the maximum real-time sample rate, which is the same across all
MSO/DPO2000/B series models.
Group
Acquisition
Syntax
ACQuire:MAXSamplerate?
Examples
ACQUIRE:MAXSAMPLERATE? might return 1.0000E+9 in a DPO2024 instrument
indicating that the maximum real-time sample rate is 1 GS/s.
ACQuire:MODe
Sets or returns the acquisition mode of the oscilloscope for all live waveforms.
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-59