EasyManuals Logo
Home>Tektronix>Test Equipment>DPO2014

Tektronix DPO2014 User Manual

Tektronix DPO2014
272 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #138 background imageLoading...
Page #138 background image
Acquire the Signal
Using FilterVu
By default, the noise lter cutoff frequency is set to the full bandwidth of the oscilloscope when you use FilterVu. To invoke the
variable low-pass lter (noise lter), turn th e multipurpose knob counterclockwise. The oscilloscope passes lower frequency signals
which results in a cleaner-looking waveform. The noise frequency readout displays frequencies that the oscilloscope includes in
the foreground waveform. T he oscilloscope displays higher frequency components in the background waveform, up to the
bandwidth of the DPO2000 or MSO2000 model.
NOTE. The oscilloscope always displays the noise lter cutoff f reque ncy in the upper right part of the screen. The cutoff
frequency reects the -3 dB point.
1. Push the FilterVu button.
120 DPO2000 and MSO2000 S eries Oscilloscopes User Manual

Table of Contents

Other manuals for Tektronix DPO2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO2014 and is the answer not in the manual?

Tektronix DPO2014 Specifications

General IconGeneral
BrandTektronix
ModelDPO2014
CategoryTest Equipment
LanguageEnglish

Related product manuals