Humidity:
Operating: 5% to 95% relative humidity (RH) at up to +30 °C
Operating: 5% to 45% relative humidity (RH) above +30 °C up to + 50 °C, non-condensing, and as limited by a Maximum Wet-Bulb
Temperature of +38 °C (derates relative humidity to 45 % RH at +50 °C)
Non-operating: 5% to 95% Relative Humidity (RH) at up to +30 °C
Non-operating: 5% to 45% Relative Humidity (RH) above +30 °C up to +50 °C, non-condensing, and as limited by a Maximum
Wet-Bulb Temperature of +38 °C (derates relative humidity to 27% RH at 60 C)
Altitude:
Operating: 3,000 m (9,8 43 ft)
Nonoperating Altitude: 12,000 m (39,370 ft)
Random Vibration:
Operating: 0.31 G
RMS
, 5 500 Hz, 10 minutes per axis, 3 axes (30 minutes total)
Non-operating: 2.46 G
RMS
, 5 500 Hz, 10 minutes per axis, 3 axes (30 minutes total)
Pollution Degree: 2, I ndoor use only
Acquisition System: 1 M
The maximum input vo ltage: At front p anel connector, 300 V
RMS
, Installation Category II; for measurements performed on circuits
directly connected to the low-voltage installation.
Acquisition System: 50 and 75
The maximum input voltage : 5 V
RMS
with a peak at ±20. For Installation Category I measurements. Not for co nnection to Installation
Category II, III, or IV circuits.
DPO3000 Series Oscilloscopes User Manua l 9

Table of Contents

Other manuals for Tektronix DPO3034

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO3034 and is the answer not in the manual?

Tektronix DPO3034 Specifications

General IconGeneral
BrandTektronix
ModelDPO3034
CategoryTest Equipment
LanguageEnglish

Related product manuals