Appendix: Warranted Specifications, Safety Certi fications, and Electromagnetic Compatibility
Appendix: Warranted Specifications, Safety Certifications, and
Electromagnetic Compatibility
The limits stated b elow are for ambient temperature of ≤30 °C and the bandwidth selection set to FULL.
Reduce the upper bandwidth frequency by 1% for each °C abo ve 30 °C.
Instrument
10 mV/div to 1 V/div 5 mV/div to 9.98 mV/div
DPO3054 and
DPO3052
DC to 500 MHz DC to 400 MHz
Instrument
2 mV/d iv to 4.9 8 mV/div 1 mV/div to 1.99 mV/div
DPO3054 and
DPO3052
DC to 250 MHz DC to 150 MHz
Instrument
5 m V /div to 1 V/div 2 mV to 4.98 mV/div 1 mV/div to
1.99 mV/div
DPO3034 and
DPO3032
DC to 300 MHz DC to 250 MHz DC to 150 MHz
Instrument
1mV/divto1V/div
Analog bandwidth,
50 Ω
DPO3014 and
DPO3012
DC to 100 MHz
DPO3000 Series Oscilloscopes User Manua l 273