Trigger Setup
Push a side-bezel button to select sequencing
the B trigger after the A as Time or as Events.
Time
(a) 8 ns
BEvents
1
Set to
Minimum
5. Set the other Sequence Trigger para m eters in
the related side- and lower-bezel menus.
B Trigger After Delay Time
The A trigger arms the instrument. Posttrigger
acquisition starts on the rst B edge after the
trigger delay time.