EasyManuals Logo
Home>Tektronix>Measuring Instruments>P6021

Tektronix P6021 User Manual

Tektronix P6021
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #38 background imageLoading...
Page #38 background image
Service Information
28
P6021 Instruction Manual
8. Remove the two Posidriv screws that secure the spring retainer holder to the
probe body; then remove the small black plastic holder.
9. Lift the circuit board, transformer, and cable out of the probe body as a unit.
If you need to, you can unplug the transformer from the circuit board.
10. To reassemble the probe, reverse the procedure above.
When replacing the slide switch, spring retainer, and transformer top as a
unit, push the slide switch contacts gently inside the sides of the bottom
housing.
To make repairs inside the probe body, disassemble the probe as described in the
previous section. If you need to solder on the circuit board, use a minimum of
heat, and observe normal circuit board procedures.
If you need to replace the current transformer, replace the entire assembly
including the other half of the transformer core mounted in the slide switch. The
transformer halves are matched at the factory before shipment.
Repairing the termination can consist of replacing either the connectors or the
circuit board. These tasks are described below.
Replacing the Connectors. To replace the connectors, follow these steps.
1. Insert a small screwdriver between the cover and the termination near the
part number, and gently pry up the plastic snap-on cover from the termina-
tion.
2. Using a heat sink, unsolder the leads from the defective connector.
3. Unscrew and remove the defective connector.
4. Replace the defective connector with the new one.
5. Screw and solder the new connector back in place.
6. Align the switch with the slider in the front cover, and replace the front
cover.
Replacing the Circuit Board. To replace the circuit board, follow these steps.
1. Remove the plastic snap-on covers from the front and back of the termina-
tion.
2. Using a heat sink, unsolder the leads from the connectors.
3. Unscrew the two screws from the back of the circuit board.
Repairing the Probe
Repairing the Termination

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix P6021 and is the answer not in the manual?

Tektronix P6021 Specifications

General IconGeneral
BrandTektronix
ModelP6021
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals