Service Information
P6021 Instruction Manual
27
retainer
Spring
Spring
Upper half of
probe body
Ball bearing
Slide switch
Pozidriv screws (2X)
holder
Spring retainer
Circuit board
Top half of
transformer
Bottom half of
transformer
Probe nose
Bottom housing
Strain relief boot
Cable
Detent
Figure 9: Disassembling the probe
1. Hold the probe horizontally with the slide switch up.
2. Move the strain relief boot back over the cable.
3. Be careful not to lose the small ball bearing you are about to expose.
4. Slowly lift the upper half of the probe body slightly at the cable end, and
push the assembly forward over the nose and off.
5. Remove the small ball bearing from the detent in the slide switch.
6. Lift the back of the return spring retainer out of the holder.
7. Remove the slide switch, spring retainer, and the top of the transformer as a
unit. Note the orientation of the movable portion of the transformer in the
slide.