e
.
ADJUST-C4205
(see
Fig
.
4-4)
for
best
square
front
corner
(see
Fig
.
4-5
for
example)
.
Disconnect
test
equip-
ment
.
f
.
Set
CH
2
VOLTS/DIV
to
10
mV
and
X1-X10
to
X1
(fully
ccw)
.
g
.
Connect
a
10X
probe
to
the
CH
2
input
.
Connect
the
probe
tip
to
a
probe
tip-to-BNC
adapter,
the
adapter
to
a
50
O
BNC
termination,
and
the
termination
to
a 50
O
BNC
10X
attenuator
attached
to
the
square-wave
generator
high-amplitude
output
connector
.
Set
generator
for
a
5-
division,
1
kHz
display
.
h
.
Compensate
probe
for
best
front
corner
of
waveform
.
REV
.
A,
JUNE
1976
m
.
Disconnect
test
equipment
.
Adjustments-T921
/T922/T922R
i .
Set
CH
2
VOLTS/DIV
to
20
mV
and
set
generator
fora
5-division
display
(remove
10X
attenuator
if
necessary)
.
j
.
ADJUST-C4213
for
flat
top
on
square
wave
.
k
.
Set
CH
2
VOLTS/DIV
to
.2
V,
and
square-wave
generator
output
for
a
5-division
display
(remove
10X
attenuator,
and
also
50
S2
termination
if
necessary)
.
I .
ADJUST-C4204
for
a
flat
top
on
square
wave
.
4-
9