TAS 200 Series Specifications
6
Bench Test Instruments and Handheld Oscilloscopes Basic Service
Table 8: Environmental characteristics
Characteristic Description
Temperature
Operating 0° to 40° C
Nonoperating -- 4 0 ° to +70° C
Tested to MIL-T-28800D, paragraphs 4. 5.5.1.3 and 4.5.5.1.4, except in 4.5.5.1.3 steps
4and5(0° C operating test) are performed ahead of step 2 (--40° C nonoperating
test). Equi pment shall remain off upon return to room ambient during step 6.
Excessive condensati on shall be removed before operating during step 7.
Altitude
Operating ≤4,570 met ers (15,000 feet). Maximum operating temperature decreases 1° C per
300 m (1,000 feet) above 1,500 m (5,000 feet).
Nonoperating ≤15,250 meters (50,000 f eet)
Relative Humidity
Operating 30° to 40° C, 90% relative humidity +0%, -- 5%
Nonoperating 30° to 60° C, 90% relative humidity +0%, --5%
Five cycles (120 hours) referenced to MIL-T-28800D paragraph 4.5.1.2.2 for type III,
class 5 instrument s. Operating and nonoperating at 90%, +0%, --5% relative humi dity.
Vibration (Operating) 15 minut es along each of three major axes at a total displ acement of 0.015 inch
peak-to-peak (2.4 g at 55 Hz) with frequency varied from 10 Hz to 55 Hz to 10 Hz in
one minut e sweeps. Hold for 10 minutes at 55 Hz in each of three major axes. All
major resonances must be above 55 Hz.
Shock (Operati ng and Nonoperating) 30 g, half-sine, 11 ms duration, three shocks per axis each direction, for a total of 18
shocks.
Radiated and Conducted Emissions Meets EN55011, class A.
Safety Listed CSA C22.2 No. 231 and UL1244
Table 9: Typical mechanical characteristics TAS 200 series
Characteristic Description
Dimensions 32.7 cm × 16.2 cm × 46.3 cm (12.9 inch × 6.38 inch × 18.2 inch)
Weight 7.0 kg (15.4 lbs)
Operating Voltages 100 V, 120 V, 220 V, 240 V,
10% variati on f rom selector range permitted.
Line Frequency 50 or 60 Hz
Power Consumption 70 VA