EasyManuals Logo
Home>Tektronix>Test Equipment>TAS 200 Series

Tektronix TAS 200 Series User Manual

Tektronix TAS 200 Series
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #33 background imageLoading...
Page #33 background image
TAS 200 Series Adjustment Procedures
Bench Test Instruments and Handheld Oscilloscopes Basic Service
29
3. Rotate the front panel HORIZONTAL POSITION control to move the dot to
the display center. Rotate the INTENSITY control to give the dot a sharp
edge.
4. Adjust VR604 on the Power and High Voltage board and the oscilloscope
front panel FOC US control until the dot becomes circular. See Figure 7 for
the adjustment location.
5. Set the front panel HORIZONTAL X-Y push switch to the out position.
6. Repeat steps 1 through 5 for best performance.

Other manuals for Tektronix TAS 200 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TAS 200 Series and is the answer not in the manual?

Tektronix TAS 200 Series Specifications

General IconGeneral
BrandTektronix
ModelTAS 200 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals