EasyManuals Logo
Home>Tektronix>Amplifier>TCPA300

Tektronix TCPA300 User Manual

Tektronix TCPA300
82 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
Troubleshooting and Error Codes
Table 6: Troubleshooting (cont.)
Problem Remedy
Degauss takes longer than 10 seconds
Probe is attached to an energized circuit. Disconnect probe from circuit and retry.
Probe is faulty - Probe transformer (defective Hall device with excessive noise or drift)
may cause this symptom. May also be caused by a shorted or open wire in the probe
cable assembly.
Defective main board - If the probe is not the cause, then it is most likely a defective
main board in the amplier.
Cannot make a current measurement -
no measurement output from amplier
Current P robe is not locked. Lock the current probe.
Current Probe is improperly connected to the amplier input connector.
Amplier COUPLING is set to AC. Set the COUPLING to DC.
A Degauss/Autobalance routine has not been completed successfully on the system.
Degauss the probe.
Oscilloscope/Amplier is not set to an appropriate sensitivity setting.
Defective interface cable (between oscilloscope and amplier).
MANUAL BALANCE will not adjust Oscilloscope or amplier input coupling not set to DC. MANUAL BALANCE buttons only
function when the ampli er is in the DC coupling m ode.
Oscilloscope/Amplier is not set to an appropriate sensitivity setting.
Stray DC component in measurement Degauss the probe using the PROBE DEGAUSS AUTOBALANCE button. (The Current
Probe has been overloaded, dropped, or exposed to magnetic eld.)
You can use the MANUAL BALANCE buttons to null out any residual DC offset after
performing a Probe Degauss Autobalance routine.
Measurements are inaccurate Degauss the probe.
The amplier output is not terminated into 50 load. Set the input impedance of
oscilloscope to 50 or connect a 50 feedthrough termination at the oscilloscope input.
Do not attach the termination at the amplier output.
The measurement exceeds the maximum continuous current or Ampere-second product
ratings of the Current Probe.
Amplier or current probe out of calibration. (See the service manual for performance
verication procedures.)
Defective current probe transformer.
Measurements roll off at high
frequencies
The oscilloscope bandwidth limit is turned on. Verify that the bandwidth limit switch on
the oscilloscope is set to the desired bandwidth position.
(Be careful not to exceed the frequency limit of the probe used. Frequencies above the
probe design may cause overheating and damage to the probe.)
Measurements exhibit excessive noise
Current P robe is not locked. Lock the current probe.
Current Probe is improperly connected to the amplier input.
Amplier output is not terminated into 50 load. Set input impedance of oscilloscope
to 50 or connect a 50 feedthrough termination at the oscilloscope input, not at the
amplier output.
Defective current probe transformer.
TCPA300/400 Ampliers and TCP300A/400 Series Current Probes User Manual 43

Table of Contents

Other manuals for Tektronix TCPA300

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TCPA300 and is the answer not in the manual?

Tektronix TCPA300 Specifications

General IconGeneral
BrandTektronix
ModelTCPA300
CategoryAmplifier
LanguageEnglish

Related product manuals