Troubleshooting and Error Codes
For example, af
ter you press the PROBE DEGAUSS AUTOBALANCE button, if
the AC and DC Coupling LEDs are ashing, then an error code is being displayed:
1. In this exampl
e, the NOT TERMINATED INTO 50 LED is illuminated.
2. This four-bit binary code (0010) indicates an Error Code 2: a null error in the
DC offset ci
rcuit occurred. (See Figure 27.)
Figure 27: Interpreting the error code display
A complete list of error codes for the amplier is listed in the table. (See Table 7.)
Table 7: Amplier error codes
Code Descrip
tion of error
Action t
o take
1
The am p
lier detected
a probe, but a valid Hall
device was not detected.
Discon
nect and reattach the probe to the amplier.
Power cycle the ampli er. Check probe transformer
(defective Hall device). If these steps do not resolve the
error,
the amplier needs service. May also be caused
by a shorted or open wire in the probe cable assembly.
2
3
An err
or occurred while
nulling out the internal DC
offset of the ampli er.
Power cycle the amplier and run the
Degauss/Autobalance adjustment routine again. If the
erro
r reoccurs, then the amplier needs service.
4
5
TCPA
300 ONLY - An
error occurred while the
amplier was adjusting
the
power supply levels
for the different Range
settings.
Powe
rcycletheampli er and run the
Degauss/Autobalance adjustment routine again. If the
error reoccurs, then perform the following steps:
1) R
emove the probe from the circuit.
2) Check probe - The signal path through the probe
transformer may be open or shorted, which can
be c
aused by defective probe transformer or cable
assembly. If the above steps do not resolve the error,
the amplier needs service.
6
7
An error occurred
while the amplier was
ad
justing the Overload
trip points.
Power cycle the amplier and run the
De
gauss/Autobalance adjustment routine again. If the
error reoccurs, then remove the probe from the circuit.
If this does not resolve the error, the amplier needs
s
ervice.
8
9
Unused
46 TCPA300/400 Ampliers and TCP300A/400 Series Current Probes User Manual

Table of Contents

Other manuals for Tektronix TCPA300

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TCPA300 and is the answer not in the manual?

Tektronix TCPA300 Specifications

General IconGeneral
BrandTektronix
ModelTCPA300
CategoryAmplifier
LanguageEnglish

Related product manuals