Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–155
HORizontal:SECdiv
Sets the time per division for the main time base and is identical to the HORi-
zontal:MAIn:SCAle command. It is included here for compatibility purposes.
Horizontal
HORizontal:SECdiv <NR3>
HORizontal:SECdiv?
:
SECdiv
?
<Space> <NR3>
HORizontal
HORizontal:TRIGger? (Query Only)
Returns the horizontal trigger parameter.
Horizontal
HORizontal:TRIGger?
:
TRIGger
?
HORizontal
HORIZONTAL:TRIGGER?
might return :HORIZONTAL:TRIGGER:POSITION 50.
HORizontal:TRIGger:POSition
Sets or queries the position of the trigger. This is equivalent to setting Trigger
Position in the Horizontal menu.
Horizontal
HORizontal:TRIGger:POSition <NR1>
HORizontal:TRIGger:POSition?