Status and Events
TDS Family Oscilloscope Programmer Manual
3–7
When output is sent to the Output Queue, the MAV bit in the SBR is set to
one (5).
When a bit in the SBR is set to one and the corresponding bit in the SRER is
enabled (6), the MSS bit in the SBR is set to one and a service request is
generated (7).
Synchronization Methods
Although most GPIB commands are completed almost immediately after being
received by the digitizing oscilloscope, some commands start a process that
requires more time. For example, once a HARDCOPY START command is
executed it may be a few seconds before the hardcopy operation is complete.
Rather than remain idle while the operation is in process, the digitizing
oscilloscope will continue processing other commands. This means that some
operations will not be completed in the order that they were sent.
Sometimes the result of an operation depends on the result of an earlier
operation. A first operation must complete before the next one gets processed.
The digitizing oscilloscope status and event reporting system provide ways to do
this.
For example, a typical application might involve acquiring a single-sequence
waveform and then taking a measurement on the acquired waveform. You could
use the following command sequence to do this:
/** Set up single-sequence acquisition **/
 
 
 
 
/** Acquire waveform data **/
 
/** Set up the measurement parameters **/
 
 
/** Take amplitude measurement on acquired data **/

The acquisition of the waveform requires extended processing time. It may not
finish before the digitizing oscilloscope takes an amplitude measurement (See
Figure 3–7). This can result in an incorrect amplitude value.

Table of Contents

Other manuals for Tektronix TDS 700A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 700A and is the answer not in the manual?

Tektronix TDS 700A Specifications

General IconGeneral
BrandTektronix
ModelTDS 700A
CategoryTest Equipment
LanguageEnglish

Related product manuals