Specications
Table1-2:Timebasesystem(cont.)
Characteristic
Description
This is the total number of samples contained in a single acquired waveform record (Memory Length in
IEEE 1057.2.2.1).
Record Length
2,500 samples per record.
5 ns/div to 50 s/div in 1–2.5–5 sequenceSeconds/Division Range
Sec/Div Variable function is not available for this product.
This is the maximum, total, long-term error in sample-rate or horizontal position time accuracy, expressed
in parts per million.
Long-Term Sample
Rate and Horizontal
Position Time Accuracy
±50 ppm over any 1 ms interval.
Horizontal scale setting Horizontal position time range
5ns/divto10ns/div
25 ns/div to 100 μs/div
250 μs/divto10s/div
25 s/div to 50 s/div
-4 div * s/div to 20 ms
-4 div * s/div to 50 ms
-4 div * s/div to 50 s
-4 div * s/div to 250 s
Horizontal Position Time
Range
The user controls the time from the trigger to the center graticule on the display with the Horizontal
Position knob.
The resolution of the Horizontal Position time is 1/25 of a horizontal division.
Horizontal Window
Display
The Window Zone enables a user to select a part of the display to be magnied. The user controls the
location with the Horizontal Position knob and the width with the Sec/Div knob. Vertical cursors show the
selected zone. To magnify the zone, the user selects the window from the menu.
This is the accuracy of delta time m easurements made on any single waveform.
The limits are given in the following table for signals having an amplitude 5 divisions, a slew rate at the
measurement points of 2.0 divisions/ns, and acquired 10 mV/div.
Condition
Time Measurement Accuracy
Single shot, sample mode, full bandwidth selected ±(1 Sample Internal + 100 ppm * |reading| +
0.6 ns)
> 16 averages, full bandwidth selected ±(1 Sample Internal + 100 ppm * |reading| +
0.4 ns)
Delta Time Measurement
Accuracy
The Sample Interval is the time between the samples in the waveform record.
Time Measurement
Functions
Frequency, Period, Rise Time, Fall Time, Positive Width, negative Width, Positive Duty Cycle, Phase,
Delay
1–6 TDS2000C Series Oscilloscope Service Manual

Table of Contents

Other manuals for Tektronix TDS2001C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS2001C and is the answer not in the manual?

Tektronix TDS2001C Specifications

General IconGeneral
BrandTektronix
ModelTDS2001C
CategoryOscilloscope
LanguageEnglish

Related product manuals