Specications
Table 1-1: Signal acquisition system characteristics (cont.)
With the same signal applied to each channel, CMRR is the ratio of the acquired signal amplitude to the
amplitude of the MAT H difference waveform, either (Channel 1 - Channel 2), (Channel 2 - Channel 1),
(Channel 3 - Channel 4) or (Channel 4 - Channel 3)
Model
Common Mode Rejection R atio
TDS2012C
TDS2014C
TDS2022C
TDS2024C
100:1 at 60 Hz, reducing to 10:1 with 50 MHz sine wave, with equal
Volts/Div and Coupling settings on each channel.
Common Mode Rejection
Ratio (CMRR), typical
TDS2001C
TDS2002C
TDS2004C
100:1 at 60 Hz, reducing to 20:1 with a sine wave with frequency equal to
½ the -3 dB bandwidth and with equal Volts//Div and Coupling settings
on each channel.
Section 4.11.1 of IEEE std. 1057. It is the ratio of the level of a signal input into one channel to that of
the s ame signal present in another channel due to stray coupling.
Model
Crosstalk
TDS2022C
TDS2024C
100:1 with a 100 MHz sine wave and with equal V/div settings on each
channel
TDS2012C
TDS2014C
100:1 with a 50 MHz sine wave and with equal V/div settings on each
channel
TDS2002C
TDS2004C
100:1 with a 30 MHz sine wave and with equal V/div settings on each
channel
Crosstalk (Channel
Isolation)
TDS2001C 100:1 with a 20 MHz sine wave and with equal V/div settings on each
channel
Time Base System
Table 1-2: Time base system
Characteristic
Description
This is the range of real-time rates, expressed in samples/second, at which a digitizer samples signals a t
its inputs and stores the samples in memory to produce a record of time-sequential samples. (IEEE
1057, 2.2.1)
Model
Sample-rate range
TDS2012C
TDS2014C
TDS2022C
TDS2024C
5 S/s to 2000 MS/s
TDS2002C
TDS2004C
5 S/s to 1000 MS/s
Sample-Rate Range
TDS2001C 5 S/s to 500 MS/s
(Sin x)/x interpolationWaveform Interpolation
Waveform interpolation is activated for sweep speeds of 100ns/div and faster.
TDS2000C Series Oscilloscope Service Manual 1–5

Table of Contents

Other manuals for Tektronix TDS2001C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS2001C and is the answer not in the manual?

Tektronix TDS2001C Specifications

General IconGeneral
BrandTektronix
ModelTDS2001C
CategoryOscilloscope
LanguageEnglish

Related product manuals