Reference
112
TDS1000/2000-Series Digital Oscilloscope User Manual
Vertical
You can use the vertical controls to display waveforms, adjust
vertical scale and positi on, and set input parameters. Refer to
page 93 for the vertical math de scriptions.
Channel Vertical Menus
There is a separate vertical menu for each channel. Each option is set
individually for each channel.
Options Settings Comments
Coupling DC
AC
GND
DC passes both AC and DC components of
the input signal
AC blocks the DC component of the input
signal and attenuates signals below 10 Hz
GND disconnects the input signal
BW Limit 20 MHz*
Off
Limits the bandwidth to reduce display
noise; filters the signal to reduce noise and
other unwanted high frequency components
Volts/Div Coarse
Fine
Selects the resolution of the Volts/Div knob
Coarse defines a 1-2-5 sequence. Fine
changes the resolution to small steps
between the coarse settings
Probe 1X
10X
100X
1000X
Settomatchthetypeofprobeyouareusing
to ensure correct vertical readouts
Invert On
Off
Inverts the waveform
* Bandwidth is reduced to 7 MHz with a 1X probe.

Table of Contents

Other manuals for Tektronix TDS2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS2002 and is the answer not in the manual?

Tektronix TDS2002 Specifications

General IconGeneral
BrandTektronix
ModelTDS2002
CategoryTest Equipment
LanguageEnglish

Related product manuals