Reference
TDS1000/2000-Series Digital Oscilloscope User Manual
79
Autoset
When you push the AUTOSET button, the oscilloscope identifies the
type of waveform and adjusts controls to produce a usable display of
the input signal.
Function Setting
Acquire mode Adjusted to Sample or Peak Detect
Display format Set to YT
Display type Set to Dots for a video signal, set to Vectors
for an FFT spectrum; otherwise, unchanged
Horizontal position Adjusted
SEC/DIV Adjusted
Trigger coupling Adjusted to DC, Noise Reject, or HF Reject
Trigger holdoff Minimum
Trigger level Set to 50%
Trigger mode Auto
Trigger source Adjusted; refer to page 80; cannot use Autoset
on the EXT TRIG signal
Trigger slope Adjusted
Trigger type Edge or Video
Trigger Video Sync Adjusted
Trigger Video Standard Adjusted
Vertical bandwidth Full
Vertical coupling DC (if GND was previously selected); AC for a
video signal; otherwise, unchanged
VOLTS/DIV Adjusted