Reference
78
TDS1000/2000-Series Digital Oscilloscope User Manual
Acquisition mode SINGLE SEQ button
Sample, Peak Detect Sequence is complete when one acquisition is acquired
Average Sequence is complete when the defined number of
acquisitions is reached (refer to page 74)
Scan Mode Display. You can use the Horizontal Scan acquisition mode
(also called Roll mode) to continuously moni tor signals that change
slowly. The oscilloscope displays waveform updates from the left to
the right of the screen and erases old points as it displays new points.
A moving, one-division-wide blank section of the screen sepa rates
the new waveform points from the old.
The oscilloscope changes to the Scan acquisition mode when you
turn the SEC/DIV knob to 100 ms/div or slower, and select the Auto
Mode option in t he TR IGGER Menu.
To disable Scan mode, push the TRIG MENU button and set the
Mode option to Norma l.
Stopping the Acquisition. While the acquisition is running, the
waveform display is live. Stopping t he acquisition (when you push
the RUN/STOP button) freezes the display. In either mode, the
waveform display can be sca led or positioned with the vertical a nd
horizontal controls.

Table of Contents

Other manuals for Tektronix TDS2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS2002 and is the answer not in the manual?

Tektronix TDS2002 Specifications

General IconGeneral
BrandTektronix
ModelTDS2002
CategoryTest Equipment
LanguageEnglish

Related product manuals