Specifications
These speci fications apply to a ll TDS2000C and TDS1000C-EDU series
oscilloscopes. To verify that a n oscilloscope meets specifications, it must first
meet the foll
owing conditions:
The oscilloscope must have been operating continuously for twenty minutes
within the s
pecified operating temperature.
You must perform the Do Self Cal operation, accessible through the Utility
menu, if th
e operating temperature has changed by more than 5 °C (9 °F)
since the last tim e the Do Self Cal operation was performed.
Specifica
tions are provided in the following tables. All specifications are
guaranteed unless noted "typical." Specifications that are marked with the symbol
are checked in the Performance Verification section. (See page 27, Performance
Verification.)
Signal Acquisition System Characteristics
NOTE.
All amplitude-related or modified specifications require 1X probe
attenuation factors unless otherwise specified. This is due to the way the d isplayed
sensitivity works. This does not affect actual methods of attachment. It only alters
the relationship between displayed scale factors and the specifications.
Table 1: Signal acquisition system characteristics
Characteristic
Description
TDS1001C-EDU
TD
S1002C-EDU
TDS1012C-EDU
TDS2001C
TD
S2002C
TDS2012C
TDS2022C
Two nominally identical channels, sampled
simultaneously.
Number of Input
Ch
annels
TDS2004C
T
DS2014C
TDS2024C
Four nominally identical channels, sampled
simultaneously.
Input Coupling DC, AC, or GND
Input Impedance, DC
Coupled
1MΩ ±2%inparallelwith20pF±3pF
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual 1