Specications
Time Base Syst
em
Table 2: Time base system
Characteris
tic
Description
This is the ra
nge of real-time rates, expressed in samples/second, at which a digitizer samples signals at
its inputs and stores the samples in memory to produce a record of time-sequential samples. (IEEE
1057, 2.2.1)
Model
Sample-rate range
TDS2012C
TDS2014C
TDS2022C
TDS2024C
5 S/s to 2000 MS/s
Sample-Rate
Range
TDS1001C-EDU
TDS1002C
-EDU
TDS1012C-EDU
TDS2001C
TDS2002
C
TDS2004C
5 S/s to 1000 MS/s
(Sin x)/x interpolationWaveform Interpolation
Waveform interpolation is activated for sweep speeds of 100 ns/div and faster.
This is the total number of samples contained in a single acquired waveform record (Memory Length in
IEEE 1057.2.2.1).
Recor
d Length
2,500 samples per record.
5 ns/div to 5 0 s/div in 1–2.5–5 sequenceSeconds/Division Range
Sec/Div Variable function is not available for this product.
Thi
s is the maximum, total, long-term error in sample-rate or horizontal position time accuracy, expressed
in parts per million.
Long-Term Sample
Rate and Horizontal
Po
sition Time A ccuracy
±5
0 ppm over any 1 ms interval.
Horizontal scale setting Horizontal position time range
5ns/divto10ns/div
2
5 ns/div to 100 μs/div
250 μs/divto10s/div
25 s/div to 50 s/div
-4 div * s/div to 20 ms
-
4 div * s/div to 50 ms
-4 div * s/div to 50 s
-4 div * s/div to 250 s
Horizontal Position Time
Ra
nge
The user controls the time from the trigger to the center graticule on the display with the Horizontal
Position knob.
The resolution of the Horizontal Position time is 1/25 of a horizontal division.
Horizontal Window
Display
The Window Zone enables a user to select a part of the display to be m agnied. The user controls the
location with the Horizontal Position knob and the width with the Sec/Div knob. Vertical cursors show the
selected zone. To magnify the zone, the user selects the window from the menu.
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual 7

Table of Contents

Other manuals for Tektronix TDS2002C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS2002C and is the answer not in the manual?

Tektronix TDS2002C Specifications

General IconGeneral
BrandTektronix
ModelTDS2002C
CategoryOscilloscope
LanguageEnglish

Related product manuals