Specications
Table2:Timebasesystem(cont.)
Characteristic
Description
This is the accuracy of delta time measurements made on any single waveform.
The limits are given in the followin g table for signals having an amplitude 5 divisions, a slew rate at the
measurement points of 2.0 divisions/ns, and acquired 10 mV/div.
Condition
Time Measurement Accuracy
Single shot, sample mode,
full bandwidth selected
±(1 Sample Internal + 100 ppm * |reading| + 0.6 ns)
> 16 averages,
full bandwidth selected
±(1 Sample Internal + 100 ppm * |reading| + 0.4 ns)
Delta Time Measurement
Accuracy
The Sample Interval is the time between the samples in the waveform record.
Time Measurement
Functions
Frequency, Period, Rise Time, Fall Time, Positive Width, negative Width, Positive Duty Cycle, Phase,
Delay