Specications
Table 1: Signal acquisition system characteristics (cont.)
Characteristic
Description
This is analog bandwidth when the instrument is DC coupled in sample or average mode. V/div values
are a ccurate for probe attenuation settings of 1X. No probe should be installed for these measurements.
System bandwidth is type tested to be equivalent to this specication with the provided probe in 10X
mode. Use Section 4.6.1 of IEEE 1057, with the reference frequency of 1 kHz at an amplitude of
5 divisions, driven from a 50 source with external termination at the input BNC (25 effective source).
TDS2022C, TDS2024C:
to 200 MHz for 5 mV/div through 5 V/div settings with bandw idth limit at full with temperature between
0°C and 35 °C. DC to > 160 MHz for 5 mV/div through 5 V/div settings with bandwidth limit at full for
temperatures between 0 °C and 50 °C. < 5 mV/div settings are limited to 20 MHz bandwidth.
TDS2012C, TDS2014C:
DC to 100 MHz for 5 mV/div through 5 V/div settings with bandwidth limit at full. < 5 mV/div settings
are limited to 20 MHz bandwidth.
TDS2002C, TDS2004C:
DC to 70 MHz for 5 m V/div through 5 V/div settings with the bandwidth limit at full. < 5 MV/div settings
are limited to 20 MHz bandwidth.
TDS2001C:
DC to 50 MHz for 5 m V/div through 5 V/div settings with the bandwidth limit at full. < 5 mV/div settings
are limited to 20 MHz bandwidth.
TDS1012C-EDU:
DC to 100 MHz for 5 mV/div through 5 V/div settings with bandwidth limit at full. < 5 mV/div settings
are limited to 20 MHz bandwidth.
TDS1002C-EDU:
DC to 60 MHz for 5 m V/div through 5 V/div settings with the bandwidth limit at full. < 5 MV/div settings
are limited to 20 MHz bandwidth.
Analog Bandwidth,
DC Coupled, Sample or
Average
TDS1001C-EDU:
DC to 40 MHz for 5 m V/div through 5 V/div settings with the bandwidth limit at full. < 5 mV/div settings
are limited to 20 MHz bandwidth.
This is the analog bandwidth when the instrument is DC coupled. V/div values are accurate for probe
attenuation settings of 1X . N o probe should be installed for these measurements.
TDS1012C-EDU, TDS2012C, TDS2014C, TDS2022C, TDS2024C:
DC to 75 MHz for 5 mV/div through 5 V/div settings with the bandwidth limit at full. Settings less
than 5 mV /div are limited to 20 MHz bandwidth.
TDS1002C-EDU, TDS2002C, TDS2004C:
DC to 50 MHz for 5 mV/div through 5 V/div settings with the bandwidth limit at full. Settings less
than 5 mV /div are limited to 20 MHz bandwidth.
Analog Bandwidth, DC
Coupled, Peak Detect,
typical
TDS1001C-EDU, TDS2001C:
DC to 30 MHz for 5 mV/div through 5 V/div settings with the bandwidth limit at full. Settings less
than 5 mV /div are limited to 20 MHz bandwidth.
Analog Bandwidth
Selections
20 MHz bandwidth limit ON/OFF
4 TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual

Table of Contents

Other manuals for Tektronix TDS2002C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS2002C and is the answer not in the manual?

Tektronix TDS2002C Specifications

General IconGeneral
BrandTektronix
ModelTDS2002C
CategoryOscilloscope
LanguageEnglish

Related product manuals