Specications
Triggering Sy
stem
Table 1-3: Triggering system
Characteris
tic
Description
Trigger Types Edge, Video, Pulse Width
Models Trigger source selections
TDS2001C
TDS2002C
TDS2012C
TDS2022C
Channel 1
Channel 2
External
External/5
AC Line
TDS2004C
TDS2014C
TDS2024C
Channel 1
Channel 2
Channel 3
Channel 4
External
External/5
AC Line
Trigger Source
Selection
External/5 selection attenuates the external signal by 5.
When Ba
ndwidth Limit is selected for a channel, the bandwidth of that channel’s trigger path will also be
limited. The bandwidth of the External Trigger path is not affected by the bandwidth limit.
Horizo
ntal Trigger
Position
The tri
gger position is set by the Horizontal Position knob.
500 ns
minimum to 10 s maximum
Trigger Holdoff Range
The ab
ility to set large values of Holdoff is limited by the difculty in adjusting the Holdoff at Sec/Div
settings less than 100 ms/Div. This is because Holdoff cannot be set in Scan Mode, which begins at
100 ms/div when Trigger Mode is A UTO. By adjusting Trigger Mode to NORMAL, the Scan Mode
oper
ation is turned off, and Holdoff can be adjusted at larger seconds/Div settings.
External Trigger Input
Impe
dance
1M±2
% in parallel with 20 pF ±3 p F
300
V
RMS
, Installation Category II; derate at 20 dB/decade above 100 kHz to 13 V peak AC at 3 MHz
and above
External Trigger
Max
imum Input Voltage
Based on sinusoidal or DC input signal. The maximum viewable signal while DC coupled is ±50 V offset
±5 V/div at 4 divisions, or 70 V. AC coupling allows measuring signals on a DC level up to 300 V. For
no
nsinusoidal waveforms, peak value must be less than 450 V. Excursions above 300 V should be less
than 100 ms duration and the duty factor is limited to < 44%. R MS signal level m ust be limited to 300 V. If
these values are exceeded, damage to the instrument may result.
Line Trigger
Characteristics
Line Trigger mode provides a source to synchronize the trigger with the AC line i nput.
I
nput Amplitude requirements: 85 V
AC
-265V
AC
.
Input Frequency requirements: 45 Hz - 440 Hz.
Edge Trigger
Trigger Modes Auto, Normal
TDS2000C Series Oscilloscope Service Manual 1–7

Table of Contents

Other manuals for Tektronix TDS2014C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS2014C and is the answer not in the manual?

Tektronix TDS2014C Specifications

General IconGeneral
BrandTektronix
ModelTDS2014C
CategoryOscilloscope
LanguageEnglish

Related product manuals