Maintenance
Flip Feet
You will need a p
air of needle-nose pliers to remove the flip feet.
Refer to the exploded view diagram. (See page 8-10.)
Removal.
1. Place the os
cilloscope face down on a soft surface such as an anti-static mat.
2. Grip the flip foot with the pliers as close to the notches as possible and
squeeze gen
tly on the foot.
3. Twist slightly to swing one hinge free of the slot.
4. Remove the flip foot from the case.
Installa
tion.
1. Place the oscilloscope face down on a soft surface such as an anti-static mat.
2. Hold the flip foot with the smooth side facing down.
3. Gently squeeze the foot so that the hinge on one side fits into the slot, and
twist slightly to snap into the slot on the other side.
Front-Panel Kn obs
Refer to the exploded view diagram for the 2-channel models or the 4-channel
models. (See page 8-4.) (See page 8-6.)
NOTE. Wrap the plier jaws with tape (such as electrical tape) to avoid scratching
the knob.
Rem
oval. Firmly grasp the knob (with pliers if necessary) and pull it away from
the front panel.
Installation. Align the keyed knob with the shaft and p ress the knob onto the shaft.
Power Button
Refer to the exploded view diagram for either the 2-channel or the 4-channel
models. (See page 8-4.) (See page 8-8.)
NOTE. Wrap the plier jaws with tape (such as electrical tape) to avoid scratching
the button.
Removal. Firmly grasp the button with pliers and pull it straight out of the cabinet.
Installation. Press the button onto the power button shaft until it snaps into place.
TDS2000C Series Oscilloscope Service Manual 6–9