Supported Active Probes and Adapters
The oscilloscope suppli es power to acti ve probe s. You ca n use the
following a ctive probes as long as the total power required by the
probes does not exceed the capacity of the oscilloscope. To
determine the tota l probe load, add the load factors for all probes you
want to use. The oscilloscope can supply power to this combination
if the sum of the l oad factors equals 10 or less. All passive probes
have a loa d factor of ze ro.
Active probe Recommended usage Load factor
P6205 FET probe, 750 MHz, 10X 0
P6243 SMT probe, 1 GHz, 10X 0
P5205 High-voltage differential probe,
1300 V, 100 MHz, 50X or 500X
6
P5210 High-voltage differential probe,
5600 V, 50 MHz, 100X or 1000X
6
ADA400A
Differential preamplifier, 10V sensitiv-
ity, DC to 10 kHz
5
AFTDS
Telecom differential 50 Ω adapter
0
AMT75
Telecom 75 Ω adapter
0
TCP202 Current probe, 15 A, DC to 50 MHz 4
013-0278-00 Video display clamp 5
CAUTION. To avoid a measurement error, do not connect act ive
probes with a combined load factor that is greater than 10. The
signal distortion caused by such an overload can be subtle (reduced
gain, dynamic range, or slew rate).