Installation Instructions
28
WVR6UP, WVR70UP, and WVR7UP Upgrades
Perform the following steps to complete the hardware installation:
23. Use a 5/16” open-end wrench to tighten the coaxial cable connectors firmly,
but not excessively.
24. If your instrument had an Audio board installed, perform the Install Upgrade
Audio Board procedure starting on page 13 to reinstall the board.
25. If you installed an Audio board, before replacing the top cover, check that
your instrument appears as shown below.
26. Replace the top cover.
27. After installing the Eye board, go to to page 29 and perform the steps in the
Option Activation section.
WVR6100 and W VR7100
WVR6100 and W VR7100
Option PHY Only