51
Appendix A
Specifications
Specifications are subject to change without notice.
Sensor Type 90° light scattering
Range 8530 Desktop 0.001 to 150 mg/m
3
8531 Desktop HC 0.001 to 400 mg/m
3
8532 Handheld 0.001 to 150 mg/m
3
Resolution ±0.1% of reading of 0.001 mg/m
3
, whichever is
greater
Zero Stability ±0.002 mg/m
3
24 hours at 10 sec time constant
Particle Size Range Approximately 0.1 to 10 µm
Flow Rate 3.0 L/min set at factory
1.4 to 3.0 L/min adjustable
Flow Accuracy ±5% factory setpoint
Internal flow controlled
Temperature Coefficient +0.001 mg/m
3
per °C
Operational Temp 0 to 50°C
Storage Temp -20 to 60°C
Operational Humidity 0-95% RH, non-condensing
Time Constant Adjustable 1 to 60 seconds
Data Logging 45 days at 1 minute samples
Log Interval 1 second to 1 hour
Physical Size (HWD) Handheld: 4.9 x 4.75 x 12.45 in.
Desktop: 5.3 x 8.5 x 8.8 in.
Weight Handheld: 2.9 lb, 3.3 lb with battery
Desktop: 3.45 lb, 4.45 lb 1 battery,
5.45 lb 2 batteries
Communications 8530/31: USB (Host and Device) and Ethernet.
Stored data accessible using thumb
drive
8532: USB (Host and Device). Stored dada
accessible using thumb drive.
Power—AC AC power adapter included.
115 to 240 VAC

Other manuals for TSI Instruments DUSTTRAK 8533

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TSI Instruments DUSTTRAK 8533 and is the answer not in the manual?

TSI Instruments DUSTTRAK 8533 Specifications

General IconGeneral
BrandTSI Instruments
ModelDUSTTRAK 8533
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals