ALL LOG FILE DATES ARE YEAR 2000
The data logger has a rechargeable battery to power
ash memory within the data logger. If this battery has
no charge, the date will revert to the default date of
the year 2000.
5.3.6
Possible Cause 1:
Battery not charged.
Action:
Connect the Flowmaster to the computer via the USB
cable and start up the FMS software. Power on the
Flowmaster. Leave like this for at least three hours.
To verify, power off the meter and disconnect from
your computer. Reconnect and set the date and time.
Power off and disconnect again. Reconnect and verify
the set date and time have been retained in data logger
memory.
Possible Cause 2:
Battery fault.
Action:
The battery may have developed a fault or may have
dislodged from it’s holder in the data logger. Try charging
the data logger battery as described above. If this fails,
return your meter to TSI for repair.
5.4 CALIBRATION OF YOUR FLOWMASTER
Every TSI ow meter is tested and calibrated prior to
shipping. A calibration label is attached to the top lid of
the meter (Figures 5.4 and 5.5). The calibration date and
the date on which the meter is due its next calibration
are written on the label for your convenience.
Fig. 5.4 Calibration label on lid of Flowmaster
Fig. 5.5 Calibration label showing calibration date and
re-calibration date
The master meter at TSI is calibrated annually at the
Irish National Test Centre. This ensures the readings
obtained by the Test Equipment referred to on
certicates issued by TSI Flowmeters Ltd are traceable
to National and International standards
Annual calibration of the Flowmaster is recommended,
particularly if the meter is used on a regular basis.
Please return your meter to TSI Flowmeters Ltd for
calibration.
Version 2.0 (February 2012) TSI Flowmeters Ltd.18