EasyManuals Logo
Home>TSI Instruments>Test Equipment>PORTACOUNT 8048

TSI Instruments PORTACOUNT 8048 User Manual

TSI Instruments PORTACOUNT 8048
101 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #82 background imageLoading...
Page #82 background image
80 Chapter 8
Always perform the Daily Checks as described in Chapter 4 as the
first troubleshooting step. Passing the Daily Checks usually indicates
that the PortaCount Fit Tester is working properly and that the
problem is elsewhere, such as the connection to the respirator and/or
the respirator itself.
On-line Troubleshooting Guides
Additional application notes on troubleshooting are located on the
FitPro™ Ultra Software or visit TSI’s website.
Symptom
Possible Causes
Solution
PortaCount Fit
Tester status
indicator is red
Hardware error
Prepare to return the unit
for service.
PortaCount Fit
Tester status
indicator is yellow;
during a
measurement
FitPro Ultra
displays a
message
concerning low
alcohol
Alcohol in wick is low.
Replenish alcohol
Low Particle
message
Low on alcohol.
Replenish alcohol.
Hoses are reversed on
Twin Tube Assembly.
Verify the hoses are
connected properly to the
PortaCount Fit Tester and
respirator.
Wrong tube is connected
to respirator.
Connect the proper tube
to the respirator (clear
tube).
Twin Tube Assembly is
kinked, pinched, or
blocked.
Straighten out the Twin
Tube Assembly or remove
the obstruction.
Particle count in area
really is low.
Move to another location
or use a particle generator
(such as TSI Model 8026
Particle Generator or an
ultrasonic humidifier).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TSI Instruments PORTACOUNT 8048 and is the answer not in the manual?

TSI Instruments PORTACOUNT 8048 Specifications

General IconGeneral
BrandTSI Instruments
ModelPORTACOUNT 8048
CategoryTest Equipment
LanguageEnglish

Related product manuals