Page #10 background imageLoading...
Page #10 background image
Note
If the value of voltage to be measured is unknown,
use the maximum measurement position (1000V) and
reduce the range step by step until a satisfactory
reading is obtained.
The LCD displays “1” indicating the existing selected
range is overloaded, it is required to select a higher
range in order to obtain a correct reading.
In each range, the Meter has an input impedance of
approx. 10M.This loading effect can cause
measurement errors in high impedance circuits. If the
circuit impedance is less than or equal to 10k, the
error is negligible (0.1% or less).
When DC voltage measurement has been completed,
disconnect the connection between the testing leads
and the circuit under test.
l
l
l
l
B. AC Voltage Measurement (see figure 3 with dotted line)
Warning
To avoid harms to you or damages to the Meter from
electric shock, please do not attempt to measure
voltages higher than 1000V or 750V rms although
readings may be obtained.
Take extra attention when measuring high voltages
to avoid electric shock.
To measure AC Voltage, connect the Meter as follows:
1. Insert the red test lead into the Vterminal and the
black test lead into the COM terminal.
2. Set the rotary switch to an appropriate measurement
position in V range.
3. Connect the test leads across with the object being
measured.
The measured value shows on the display, which is
effective value of sine wave (mean value response).
Model UT39A/B/C: OPERATING MANUAL
10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the UNI-T UT39A and is the answer not in the manual?

UNI-T UT39A Specifications

General IconGeneral
Diode testYes
Product typeDigital multimeter
AC current range0.0002 - 10 A
AC voltage range2 - 750 V
DC current range0.00002 - 10 A
DC voltage range0.2 - 1000 V
Capacitance range2000 nF
Basic accuracy (DC current)±(0.8%+1)
Basic accuracy (DC voltage)±(0.5%+1)
Dimensions (WxDxH)83 x 38 x 172 mm
Battery voltage9 V
Weight and Dimensions IconWeight and Dimensions
Weight310 g