4. Break the current path to be tested. Connect the red
test lead to the more positive side of the break and
the black test lead to the more negative side of the
break.
5. Turn on power to the circuit.
The measured value shows on the display.
To measure current, do the following:
1. Turn off power to the circuit. Discharge all high-voltage
capacitors.
2. Insert the red test lead into the mA or 20A or 10A
terminal and the black test lead into the COM terminal.
When measuring current at 200mA below, insert the
red test lead into mA terminal while measuring current
200mA or above, insert the red test lead into 10A or
20A terminal
3. Set the rotary switch to an appropriate measurement
position in A range.
Note
l
l
l
l
l
If the value of current to be measured is unknown, use
the maximum measurement position (20A) and 20A
terminal or (10A) and 10A terminal, and reduce the
range step by step until a satisfactory reading is
obtained.
Replace appropriate rating fuse when the fuse is burnt.
Fuse specification: 0.315A. 250V fast type fuse,
φ 5 x 20mm
UT39A/UT39B - At 10A Range: For continuous
measurement 10 seconds and interval not less than
15 minutes.
UT39C – At 20A Range: For continuous measurement
10 seconds and interval not less than 15 minutes
When current measurement has been completed,
disconnect the connection between the testing leads
and the circuit under test.
12
Model UT39A/B/C: OPERATING MANUAL