50 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
Arrhythmia
Arrhythmia settings
1. Press the Setup button .
2. Select Arrhythmia.
The default settings are in bold.
Note
The full arrhythmia menu is only displayed when the full arrhythmia option is
licensed. When the full arrhythmia option has not been enabled, settings are
made for VFib, Asystol and Vtach only.
Main menu Parameter Description
Setup Arrhythmia VFib Alarm Level - Lethal (cannot be changed), Print
on alarm Yes/No.
Asystole Alarm Level - Lethal (cannot be changed), Print
on alarm Yes/No.
VTach VTach Limit 100 to 200 (125), Alarm Level - Lethal
(cannot be changed), Print on alarm Yes/No.
PVC run PVC run 3 to 6 (6) Alarm Level (High, Medium,
Low, Off), Print on alarm Yes/No.
VRyhthm Alarm Level (High, Medium, Low, Off), Print on
alarm Yes/No.
Couplet Alarm Level (High, Medium, Low, Off), Print on
alarm Yes/No.
PVC/min PVC/min 1to 30 (30) Alarm Level (High, Medium,
Low, Off), Print on alarm Yes/No.
Bigeminy Alarm Level (High, Medium, Low, Off), Print on
alarm Yes/No.