Directions for use Monitoring and Measurements 59
etCO
2
settings
The parameter settings are selected with the trim knob, described previously (see
“Settings via a parameter field” on page 31).
The default settings are in bold.
Parameter Description
Size 0 to 40 mmHg, 0 to 60mmHg, 0 to 80mmHg.
Units mmHg/kPa.
Apnea time 6, 10, 15, 20, 25, 30 seconds.
EtCO
2
lower /upper limit Range: 0- 99 mmHg
Lower limit default 25 mmHg / 3 kPa)
Upper limit default 60 mmHg / 8 kPa.
RR lower limit / upper limit Range: 2- 150
Lower limit default 5
Upper limit default 30.
FiCO
2
upper limit 2 - 25 mmHg (5 mmHg / 0.7 kPa).