EasyManuals Logo

Welch Allyn TM286 User Manual

Welch Allyn TM286
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #67 background imageLoading...
Page #67 background image
Directions for Use Standards, Specifications, and Accessories 63
Probe LED Indicators
Audiometry mode
Transducers
Intensity Levels
Probe:
Steady yellow: Occlusion
Blinking green: Ready to start
Steady green: Test in progress
Steady orange: Leak in pressure
No Light: Test is finished
Frequencies: 125, 250, 500, 750, 1000, 1500, 2000, 3000, 4000, 6000, 8000 Hz
Accuracy: ±2%
Total Harmonic Distortion: < 2.5% (125 to 3000 Hz measured acoustically at maximum dB HL; 4000 and
6000 Hz measured electrically
Audiometric Headset Pair External Headset earphones Type 51 cushions (10 ohms impedance)
Headband force per ANSI S3.6 and IEC 645 (4.5 ±0.5)
Insert Earphones: Ear Tone Insert earphones (50 ohm impedance)
External Headset earphones
Frequency Intensity
125 Hz -10 to 50 dB HL
500 to 4000 Hz -10 to 90 dB HL
6000 Hz -10 to 85 dB HL
250 and 8000 Hz -10 to 70 dB HL
Note
An additional +10 dB is available per frequency via the +10 dB button.
Accuracy: 125 to 4000 Hz ±3 dB
6000 and 8000 Hz ±5 dB
Step Size: 5 dB
Signal-to-Noise Ratio: > 70 dB in 1/3 octave; less than -10 dB HL for levels less than 60 dB HL
Rise/Fall Time: 20 to 50 msec

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Welch Allyn TM286 and is the answer not in the manual?

Welch Allyn TM286 Specifications

General IconGeneral
BrandWelch Allyn
ModelTM286
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals