<5. Basic Operating Procedures>
9
IM 01E21A01-03EN-A
5. Basic Operating Procedures
5.1 Operation by Display unit
(Please read the descriptions on IM 01E22A01-01EN.)
5.2 Display and Basic Conguration
(Please read the descriptions on IM 01E22A01-01EN.)
5.3 Display Mode and Setting Mode
The device runs in the Display Mode when the power is turned on. For check or change of parameters, the Setting Mode
must be activated. The following procedure explains how to change to the Setting Mode. For the function of IR switches,
read Subsection 5.2.1.
[Procedure]
1) Keep touching [SET] switch for few seconds.
0.00000m/s
0.00000
l/min
00 : 00
PRV
VEL
FLB
SET SFT INC
2) Touch [ SFT] + [ IN C] s wi tch es.
Touch SFT + INC
SET SFT INC
3) “No“ is selected.
Touch [ INC] s wi tc h a nd s el ec t “ Yes”.
Setting Mode
No
Yes
▲
▼
00 : 00
SET SFT INC
4) Touch [ SET] s wi tc h.
Setting Mode
Yes
No
▲
▼
00 : 00
SET SFT INC
5) “Yes” is blinking.
Touch [ SET] s wi tc h a ga i n.
SET SFT INC
Setting Mode
Yes
6) The screen moves to the menu of Operation Level.
Operation level
Specialist
Exit
Operator
▲
▼
00 : 00
SET SFT INC
7) Select an appropreate operation level by moving the
cursor with [INC] or [DEC] switch.
Passcode is not necessary for “Operator”. For
“Maintenance” and “Specialist”, passcode is
necessary for each.
For passcode setting, [SFT] is for
position change, and [INC] is for number, then twice [SET]
is for entry completion.
The default passcode at the factory shipment is set to
“0000”.
Pass code
****
0000
0000 9999
00 : 00
SET SFT INC
8) When the Operation Level is determined, the screen
moves to “Device setup” as the Setting Mode where
parameters can be configured.
9) After completing parameter setting, push [ESC]
switch. The screen returns to the Display Mode.