15
Performing a Blood Glucose Test
Before performing a blood glucose test on patients, set up the meter correctly. The meter, a test strip, and a
disposable, single‑use lancet or blood collection device are required.
1. Wear protective gloves.
2. Wash or disinfect the patient’s nger and allow it to dry thoroughly. The blood drop does not spread so
easily over dry skin.
3. Use only a lancing device that is approved for use by healthcare professionals.
4. Use a lancing device with adjustable penetration depth settings. If using a Safe‑T‑Pro Plus Lancing Device,
select a penetration depth to suit the nature of the patient’s skin. Start with a shallow depth setting.
5. Select a puncture site along the side of a ngertip.
6. Insert a test strip into the meter.