EasyManuals Logo
Home>Accu-Chek>Blood Glucose Meter>Performa

Accu-Chek Performa Operator's Manual

Accu-Chek Performa
68 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #25 background imageLoading...
Page #25 background image
23
Chapter 3: Control Testing for Healthcare Professionals
Introduction
Testing control solutions with known glucose levels establishes that the operator and the system are
performing acceptably. Control results must be within the dened acceptable ranges before valid patient
testing is allowed.
Testing Intervals
Follow your facility’s policy for control testing intervals.
Control testing should be performed:
The rst time before using the meter for patient
testing
At intervals established by the facility
When a new box of test strips is opened
If the test strip container is left open
If test strips were incorrectly stored
If there is a question about a patient’s glucose
result
To check the performance of the system
If the meter was dropped

Other manuals for Accu-Chek Performa

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Performa and is the answer not in the manual?

Accu-Chek Performa Specifications

General IconGeneral
BrandAccu-Chek
ModelPerforma
CategoryBlood Glucose Meter
LanguageEnglish

Related product manuals