3-37
Triggering the Oscilloscope
To use SPI triggering
To use SPI triggering
Serial Peripheral Interface (SPI) trigger setup consists of connecting the
oscilloscope to a clock, data, and framing signal. You can then trigger on a data
pattern during a specific framing period. The serial data string can be specified
to be from 4 to 32 bits long.
When you press the Settings softkey, a graphic will be displayed showing the
current state of the frame signal, clock slope, number of data bits, and data bit
values. Press the Signals softkey in the Settings menu to see the current source
channels for the clock, data, and frame signals.
•Press the More key in the Trigger section of the front panel, rotate the
Entry knob until
SPI is displayed in the Trigger softkey, then press the
Settings softkey to display the SPI trigger menu.
Assign
channels
# data bits
in string
Data bit
select
Set all
data bits
to value
Trigger level
or threshold
Currently selected Clock,
Framing, or Data channel
SPI trigger
Return to
previous menu
Data bit
value
Data string
values
Graphic showing
current state of the
SPI trigger setup

Table of Contents

Other manuals for Agilent Technologies 54622A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 54622A and is the answer not in the manual?

Agilent Technologies 54622A Specifications

General IconGeneral
BrandAgilent Technologies
Model54622A
CategoryTest Equipment
LanguageEnglish

Related product manuals